[关键词]
[摘要]
简要介绍了平凉地电比测台站的场地环境,阐述了比测装置系统和观测室的建设过程,并对其性能进行了测试和分析。结果表明,建成的装置系统运行稳定,产出数据真实可靠;观测室的改建充分利用了现有基础设施,注重功能划分、防雷接地、综合布线等设计,可以满足地电比测的需求。最后,以ZD8M型地电阻率仪为例,对定型比测结果进行了分析,结果进一步说明甘肃平凉地电比测台站的各项指标均符合相关技术规范。
[Key word]
[Abstract]
In this paper we briefly introduce the site environment of Pingliang geo-electric instrument test station, the construction process of comparison device system and observation room, with which performance is tested and analyzed. The results show that the system runs stably and the output data is reliable. In the reconstruction of the observation room we make full use of the existing infrastructure, and pays attention to the design of function division, lightning protection and grounding, generic cabling and etc., which is capable of meeting the needs of geo-electric instrument comparison test. Finally, using ZD8M geo-electric resistivity meter as an example we conduct comparison and measurement analysis. The results show that the indexes of Pingliang geo-electric instrument comparison test station are in line with the relevant technical specifications.
[中图分类号]
P315
[基金项目]
甘肃省地震局科技发展野外站基金(2020Y03)资助